LBGF-8000AB--Industrial Analyzer
Other Products
Model LBGF-8000A: Tests three indicators: moisture, ash, and volatile matter in a single furnace.
Model LBGF-8000B: Tests the volatile matter indicator, designed according to the national standard method for volatile matter testing.
When LBGF-8000A and LBGF-8000B work simultaneously, the testing speed is faster, and the testing accuracy is higher. Both the rapid method and the national standard classic method are available.
Features of the High-Precision Fully Automatic Industrial Analyzer:
■ High Automation: After placing the sample, simply click the mouse to automatically complete the test and print the results. Operation is simple.
■ Accurate Testing: Tests on over twenty standard coal samples have yielded accurate results. Comparative tests with the national standard method show that the testing precision meets requirements. The LBGF-8000A and LBGF-8000B instruments adopt a dual-body structure, with volatile matter testing as a separate part of the instrument, and the other part testing moisture and ash. Test results exceed national standard requirements.
■ High Efficiency:
Rapid Method: Moisture, ash, and volatile matter for 19 samples ≤120min (LBGF-8000A/B)
≤180min (LBGF-8000A)
Classic Method: Moisture (dryness test) for 19 samples 260min, ash (slow ash + dryness test) 260min, volatile matter 120min (LBGF-8000A/B).
■ The three indicators—moisture, ash, and volatile matter—can be measured in any combination or individually.
■ Uses an imported integrated furnace, with fast heating, energy savings, and short testing times.
■ Uses imported electronic balances and imported stepper motors to ensure stable testing performance.
■ New feature: Pause the experiment midway and resume the next day. New electric push rod for automatic lid opening and closing. New wireless data sharing function allows users to wirelessly transmit data directionally.
■ Can connect to an electronic balance.
Technical Parameters of the High-Precision Fully Automatic Industrial Analyzer:
Power:
LBGF-8000A ≤5KW
LBGF-8000B ≤4KW
LBGF-8000A/B ≤9KW
Gas Requirements:
Oxygen: Purity 99.5%, pressure after减压 0.1Mpa
Nitrogen: Purity 99.5%, pressure after减压 0.1Mpa
减压器: 0-25Mpa
Low End 0-0.4Mpa
Number of Crucibles: 20, can test 19 samples at once
Sample Mass: Moisture, ash samples 0.6-1.1g (adjustable); volatile matter 0.9-1.1g
Furnace Temperature Range: Room temperature to 0-1000℃
Analysis Precision: Meets the requirements of GB/T212-2008 "Proximate Analysis of Coal" standard.
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 鹤壁蓝博 |
| Spec: | |
| Stock: | |
| Origin: | China / Henan / Hebishi |